             LTP Installation Checklist

Advia Programming

  1. Change settings for rerunning samples
  2. Select print function for Path Reviews
  3. Program additional parameters
  4. Select Alarm stop criteria
  5. Verify species settings ( 6 user defined species only)
  6. Set Patient IDEE length
  7. Set Patient Data Transmission
  8. Set Patient Review Requirement.

 

1. Change settings for rerunning samples

Changing the settings below will result in sample reruns overwriting prior results

Modify FSE parameters

Customize>SystemSet Up>Tools Modify>FSE Parameters

Enter B00MFSEA at password prompt

Select File>LOAD> wst.ini then OK

Set  Uiw_Mode = OVR

Set Uiw_Keep_RequestedTestOnly=N

Modify Customer Parameters

Customize>SystemSet Up>Tools Modify>Customer Parameters

Set item REPLACE SAMPLE IN COMPLETE WITH A WORKORDER to Y

Set Item SAMPLE SYSTEM FLAGS TRANSFER TO HOST to Y

Set Item “REVIEW NONE” MODE FOR CONTROL SAMPLES to Y

Select File>Save>OK and Exit

Check item LIST OF SAMPLE STATUS ALLOWED FOR QUERY ANSWERS is set to P/H/I/CC/C

2.  Print Function for Path reviews

In the Review Edit screen select;

Parameter, Result report format, Dedicated.

3. Program Additional Advia Parameters

The below list includes tests required for LTP use. It is likely that many will require programming. If the code is presently programmed check the test and Host numbers against the list. To check/program these go to:

Customize>System Set up>Tools Modify>Test Dictionary.

Type in a code from the following table in the open box

Select Update

Enter WB next to type (may already be prefilled)

Enter Code name next to NAME (may already be prefilled)

Enter Test number and Host number

Select OK

Code                    Test              Host

Neutx                   101               191

Neuty                   102               192

WBCB                   110               52

WBCP                   104               50

B%NO                   135              135

B%SAT                   115             115

HGBcel                  10                30

MPC                       66               66

MPM                      68              68

CLMPCT                 240            253

 

4. Change Alarm stop criteria

Go to Customize>System Set Up>Alarm/Stop Criteria

Select Sample/System flagging

Set values for Alarm and Stop criteria from list below.

NB. Any not in bold should be set to 0

Flag      Description                                              Alarm Criteria       Stop Criteria

BC         Baso Count Suspect                                        0                         0

BS         Baso Saturation                                               0                         0

TB         Baso Temperature  Out                                 0                          0

VB        Baso No Valley                                                 0                          0

NB         Baso Noise                                                       0                         0

BR          Baso Irregular Flow Rate                             1                         3       

VX          No Perox Noise/Lymph  Valley                   1                        3

NX         Perox Noise                                                      0                         0

XS          Perox Saturation                                            1                         0

TX            Perox  Temperature Out                               1                        3

PX           Perox Power Low                                            1                         3

XR           Perox Irregular Flow Rate                             1                         3

CC           Comparison Error MCHC/CHCM                    0                        0

NT           Noise Platelets                                                  0                        0

OT           Platelets Noise Origin                                     1                        0

PH            HGB Power Low                                              1                        3

PL             Laser Power Low                                             1                        3

HR            Hemoglobin Irregular Flow Rate                  1                        3

RR            RBC Irregular Flow Rate                                  1                        3

CA            Retic Absorption Distribution Abnormal      0                        0

FC             Retic Fit Suspect                                               0                        0

NO           Retic Noise Origin                                             1                       3

CT            Retic –PLT Interference/Error                         0                        0

CL            Retic RBC Count Low                                         0                        0

CS            Retic Saturation                                                 1                        3

SE            Retic Slope Error                                                 0                       0

CR           Retic Irregular Flow rate                                   1                       3

RF            Retic  Absorption Flatness                               1                       3

WC          Comparison  Error WBCP/WBCB                      0                      0

WS          WBC Substitution                                                1                      3

 

 

5. Verify Species Settings for these 6 user-defined species only

Go to Customize>System Set Up>User Defined Species

Only the six user defined species in the table below should be programmed.

Use the dropdown menu to choose the associated species.

Select Associate and Save. Enter OK at the message.

Delete any User defined Species not listed below

Highlight the species in the list to the right.

Select Delete and OK.

Name                         Associated Species

XDOG                         Dog

XCAT                          Cat

XHORSE                    Horse

Ferret                        Dog

Llama                        Horse

Marine Mammal    Dog

 

6.  Set Patient IDEE length

Go to Customize>System Set Up> Tools Modify>Format

Select File>Load>trame par and OK at warning message.

Set IDEE=(12,R,Z)

Select File>Save and Exit.

7. Set patient Data Transmission

Go to Customize>Tools View >Routine parameters.

Check the box marked Patients in the Host Transmission section

Check the box marked Controls in the Host Transmission section

 

8. Set patient Review Requirement

Go to Data Manager> Sample Control Panel>Review/Edit>Parameter.

Select Review None.

 

Note: If difficulties with data transmission arise it is worth checking the Trame file is correctly loaded.

To do this go to Customize> System Set up>Tools Modify> Format.

Select File>Load>trame par and OK at the warning message.

Scroll down to Trame_ Y and Trame_R and ensure the sequence is as shown below.

 

Please notify your QA specialist if any discrepancies with above settings arise.

 

 